Header model | XP3000 |
Detection Gas | O2 |
Detection Range | 0-30% VOL |
Detection Principle | Electrochemical |
Sensor Brand | British CITY |
Measurement Accuracy | 0.1% VOL |
Response Time | T90 < 30s |
Analog Signal Output | 4-20mA |
Power Supply | DC 8V - 30V |
Power Consumption | < 5W |
Explosion-proof Grade | ExdIICT6Gb |
Protection Grade | IP66 |
Weight | 1.5kg |
Ambient Temperature | -20°C to +50°C (Avoid direct sunlight) |
Ambient Humidity | 10% - 93% RH (Non-condensing) |
Pharmaceutical Industry:
The centrifuge reactor oxygen content analyzer is used to monitor the oxygen levels during various stages of pharmaceutical production processes.
Biochemical Field:
This analyzer plays a critical role in the research and production of biochemical technologies, ensuring precise oxygen monitoring for optimal processes.
Environmental Engineering Industry:
The analyzer is employed to monitor oxygen content in environmental engineering applications, such as the treatment of wastewater and the management of waste gases.
Gas Sampling:
A gas sampling system is employed at the laboratory or production site to extract gas from the reactor.
Gas Pretreatment:
The collected gas undergoes pretreatment to remove impurities, such as water vapor, carbon dioxide, and other contaminants.
Oxygen Sensor Detection:
The purified gas is then sent to the oxygen sensor, which detects changes in oxygen content and transmits the data to the data acquisition and processing system.
Data Acquisition and Processing:
The system collects, records, and processes the oxygen content data detected by the oxygen sensor. It generates relevant reports and enables laboratory personnel to analyze trends in oxygen content changes throughout the reaction, as well as its relationship with other process parameters.
Control System:
The data processing system analyzes and evaluates the oxygen content variations during the reaction. When required, the control system can automatically adjust oxygen levels to optimize the reaction process.